Flat panel CT versus multidetector CT in skull base imaging: are there differences in image quality?

Purpose of this study was to compare image quality of the skull base in standard 20s protocol flat panel computed tomography (FPCT) with the new time and dose improved 10s protocol as well as with 128 slice multidetector computed tomography (MDCT). 10 whole skull preparations were scanned with eith...

Description complète

Enregistré dans:
Détails bibliographiques
Auteurs principaux: Schulze, Maximilian, Hirt, Bernhard, Reimann, Katrin
Format: Article
Langue:anglais
Publié: Philipps-Universität Marburg 2024
Sujets:
Accès en ligne:Texte intégral en PDF
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
Description
Résumé:Purpose of this study was to compare image quality of the skull base in standard 20s protocol flat panel computed tomography (FPCT) with the new time and dose improved 10s protocol as well as with 128 slice multidetector computed tomography (MDCT). 10 whole skull preparations were scanned with either 128 slice MDCT(SOMATOM Definition AS+, Siemens, Erlangen) or FPCT (AXIOM-Artis, Siemens, Erlangen) using 10s or 20s protocol. FPCT provides significantly better image quality and improved delimitation of clinically relevant structures in the anterior, temporal and posterior skull base compared to 128 slice MDCT. The 20s FPCT protocol yielded best delimitability of evaluated skull base structures. However, the shorter, dose saving 10s FPCT protocol was still significantly superior to 128 slice MDCT regarding delimitability of skull base structures and additionally showed no significant inferiority compared with the 20s FPCT protocol. The 10s FPCT protocol yields a significantly better image quality at a comparable radiation dose exposure in imaging skull base structures compared to MDCT.
Description:Gefördert durch den Open-Access-Publikationsfonds der UB Marburg.
DOI:10.1186/s13005-023-00391-0