Delayed perforation of an atrial pacemaker electrode: Lifelong risk for a rare but serious complication

We report the successful percutaneous management of a 77-year- old woman who had a dual-chamber pacemaker implanted more than 1 year ago and presented with pericarditis pain and compensated pericardial hemorrhagic tamponade. The symptoms were due to very late acute perforation of the atrial le...

Cijeli opis

Spremljeno u:
Bibliografski detalji
Glavni autori: Waechter, Christian, Koenig, Alexander M., Chatzis, Georgios, Mueller, Julian, Schieffer, Bernhard, Luesebrink, Ulrich
Format: Članak
Jezik:engleski
Izdano: Philipps-Universität Marburg 2023
Teme:
Online pristup:PDF cijeli tekst
Oznake: Dodaj oznaku
Bez oznaka, Budi prvi tko označuje ovaj zapis!
Opis
Sažetak:We report the successful percutaneous management of a 77-year- old woman who had a dual-chamber pacemaker implanted more than 1 year ago and presented with pericarditis pain and compensated pericardial hemorrhagic tamponade. The symptoms were due to very late acute perforation of the atrial lead. This report is intended to raise awareness of procedure-related complications in the large group of cardiovascular implantable electronic device patients. Pleuritic or pericardial pain in these patients should raise suspicion of electrode perforation, as the risk of perforation is not restricted to the period immediately after implantation and a lifelong risk cannot apparently be excluded.
Opis predmeta:Gefördert durch den Open-Access-Publikationsfonds der UB Marburg.
Digitalni identifikator objekta:10.1002/ccr3.7525