Delayed perforation of an atrial pacemaker electrode: Lifelong risk for a rare but serious complication

We report the successful percutaneous management of a 77-year- old woman who had a dual-chamber pacemaker implanted more than 1 year ago and presented with pericarditis pain and compensated pericardial hemorrhagic tamponade. The symptoms were due to very late acute perforation of the atrial le...

Disgrifiad llawn

Wedi'i Gadw mewn:
Manylion Llyfryddiaeth
Prif Awduron: Waechter, Christian, Koenig, Alexander M., Chatzis, Georgios, Mueller, Julian, Schieffer, Bernhard, Luesebrink, Ulrich
Fformat: Erthygl
Iaith:Saesneg
Cyhoeddwyd: Philipps-Universität Marburg 2023
Pynciau:
Mynediad Ar-lein:Testun PDF llawn
Tagiau: Ychwanegu Tag
Dim Tagiau, Byddwch y cyntaf i dagio'r cofnod hwn!
Disgrifiad
Crynodeb:We report the successful percutaneous management of a 77-year- old woman who had a dual-chamber pacemaker implanted more than 1 year ago and presented with pericarditis pain and compensated pericardial hemorrhagic tamponade. The symptoms were due to very late acute perforation of the atrial lead. This report is intended to raise awareness of procedure-related complications in the large group of cardiovascular implantable electronic device patients. Pleuritic or pericardial pain in these patients should raise suspicion of electrode perforation, as the risk of perforation is not restricted to the period immediately after implantation and a lifelong risk cannot apparently be excluded.
Disgrifiad o'r Eitem:Gefördert durch den Open-Access-Publikationsfonds der UB Marburg.
DOI:10.1002/ccr3.7525